Quantitative evaluation of dosimetric uncertainties in electron therapy by measurement and calculation using the
Imad Ali1, Nesreen Alsbou2, Salahuddin Ahmad1
1Department of Radiation Oncology, University of Oklahoma Health Sciences Center, Oklahoma City, Oklahoma, USA.
Abstract:
In the electron beam radiation therapy, customized blocks are mostly used to shape treatment fields to generate conformal doses. The goal of this study is to investigate quantitatively dosimetric uncertainties associated with heterogeneities, detectors used in the measurement of the beam data commissioning, and modeling of the interactions of high energy electrons with tissue. These uncertainties were investigated both by measurements with different detectors and calculations using electron Monte Carlo algorithm (eMC) in the Eclipse treatment planning system. Dose distributions for different field sizes were calculated using eMC and measured with a multiple-diode-array detector (MapCheck2) for cone sizes ranging from 6 to 25 cm. The dose distributions were calculated using the CT images of the MapCheck2 and water-equivalent phantoms. In the umbra region (<20% isodose line), the eMC underestimated dose by a factor of 3 for high energy electron beams due to lack of consideration of bremsstrahlung emitted laterally that was not accounted by eMC in the low dose region outside the field. In the penumbra (20%-80% isodose line), the eMC overestimated dose (40%) for high energy 20 MeV electrons compared to the measured dose with small diodes in the high gradient dose region. This was mainly due to lack of consideration of volume averaging of the ion chamber used in beam data commissioning which was input to the eMC dose calculation algorithm. Large uncertainties in the CT numbers (25%) resulted from the image artifacts in the CT images of the MapCheck2 phantom due to metal artifacts. The eMC algorithm used the electron and material densities extracted from the CT numbers which resulted large dosimetric uncertainties (10%) in the material densities and corresponding stopping power ratios. The dose calculations with eMC are associated with large uncertainties particularly in penumbra and umbra regions and around heterogeneities which affect the low dose level that cover nearby normal tissue or critical structures.
Related Concept Videos
Dose Size and Dosing Frequency: Determination Methods
Propagation of Uncertainty from Systematic Error
The Uncertainty Principle
Determination of Multiple Dosing Parameters: Loading and Maintenance Doses
Dosage Regimen Designs: Nomograms and Tabulations
Propagation of Uncertainty from Random Error