Related Experiment Video
Updated: Oct 12, 2025

Setup of Capillary Electrophoresis-Inductively Coupled Plasma Mass Spectrometry CE-ICP-MS for Quantification of Iron Redox Species FeII, FeIII
Published on: May 4, 2020
Accurate quantification of phosphorus intergranular segregation in iron by STEM-EDX
C-Y Hsu1, J Stodolna2, P Todeschini2
1EDF R&D, MMC Department, F-77250 Ecuelles, France; Mines Saint-Etienne, Univ Lyon, CNRS, UMR 5307 LGF, Centre SMS, F-42023 Saint-Etienne, France.
Abstract:
This study describes a method to quantify phosphorus grain boundary segregation by Energy Dispersive X-ray Spectroscopy in Scanning Transmission Electron Microscope (STEM-EDX). A "box-type method" is employed, removing the long-discussed problems of interaction volume and the beam broadening effect. The proposed methodology also introduces a novel way of subtracting the spectrum background to remove the influence of coherent Bremsstrahlung and spurious peaks. A Fe-P model alloy was used to compare the box method to the quantification results previously obtained by atom probe tomography on two high angle grain boundaries. The results are specifically reported in surface concentration (atom/nm2) to avoid additional hypotheses and allow the results between the two techniques to be directly compared. The measurements show that the box-type method can accurately measure phosphorus intergranular segregation in iron.
More Related Videos
14:53In Situ Detection and Single Cell Quantification of Metal Oxide Nanoparticles Using Nuclear Microprobe Analysis
Published on: February 3, 2018
08:21Optimized Procedure for Determining the Adsorption of Phosphonates onto Granular Ferric Hydroxide using a Miniaturized Phosphorus Determination Method
Published on: May 18, 2018