Related Experiment Video
Updated: Oct 12, 2025

Evaluation of Patients' Posture and Gait Profile After Lumbar Fusion Surgery by Video Rasterstereography and Treadmill Gait Analysis
Published on: March 23, 2019
Test-Retest Reliability of PODOSmart® Gait Analysis Insoles
Andreas Loukovitis1, Efthymios Ziagkas1, Dimitrios Xypolias Zekakos2
1Department of Physical Education and Sport Science, Aristotle University of Thessaloniki, 57001 Thessaloniki, Greece.
Abstract:
It is recognized that gait analysis is a powerful tool used to capture human locomotion and quantify the related parameters. PODOSmart® insoles have been designed to provide accurate measurements for gait analysis. PODOSmart® insoles are lightweight, slim and cost-effective. A recent publication presented the characteristics and data concerning the validity of PODOSmart® insoles in gait analysis. In literature, there is still no evidence about the repeatability of PODOSmart® gait analysis system. Such evidence is essential in order to use this device in both research and clinical settings. The aim of the present study was to assess the repeatability of PODOSmart® system. In this context, it was hypothesized that the parameters of gait analysis captured by PODOSmart® would be repeatable. In a sample consisting of 22 healthy male adults, participants performed two walking trials on a six-meter walkway. The ICC values for 28 gait variables provided by PODOSmart® indicated good to excellent test-retest reliability, ranging from 0.802 to 0.997. The present findings confirm that PODOSmart® gait analysis insoles present excellent repeatability in gait analysis parameters. These results offer additional evidence regarding the reliability of this gait analysis tool.