Less is more: optimal recording time for measuring the steady-state accommodative response
Beatríz Redondo1, Jesús Vera1, Rubén Molina1
1Claro (Clinical and Laboratory Applications of Research in Optometry) Research Group, Department of Optics, Faculty of Sciences, University of Granada, Granada, Spain.
Abstract:
Clinical Relevance Background: Steady-state accommodation is frequently evaluated with open-field autorefractors; however there is no consensus regarding the optimal test duration. Our results allow us to standardise the procedure followed for the assessment of the steady-state accommodative response in clinical and laboratory settings and may be of great relevance in time-constrained situations.
Background:
The present study aims to determine the optimal recording time to assess the magnitude and variability of the accommodative response with a binocular open-field autorefractor.
Methods:
The steady-state accommodative response of 84 healthy undergraduate students was dynamically measured using the Grand Seiko WAM-5500 autorefractor at 500 cm, 40 cm and 20 cm for 90 s while participants maintained focus on a stationary target. We examined the reliability and level of agreement between four time intervals (the first 10-s, 30-s, 60-s and 90-s) at three accommodative demands (0.2D, 2.5D and 5D).
Results:
The magnitude of accommodation showed an excellent level of reliability (intraclass correlation coefficient [ICC]> 0.90) for the comparisons between the 90-s, 60-s and 30-s time intervals at the three viewing distances tested. However, when including the 10-s interval, there was a moderate reduction of the reliability indices and an increase of the mean difference between two time intervals (intra-class correlation [ICC] ranging from 0.89 to 0.97; and ESs ranging from 0.16 to 0.42). For the variability of accommodation, we observed an excellent reliability for the comparisons of the 90-s and 60-s measures at the three accommodative demands (ICC: 0.95-0.96), whereas for the rest of comparisons we observed from moderate to good levels of reliability.
Conclusion:
Our data indicate that of 30-s and 60-s are enough to accurately assess the magnitude and variability of the accommodative response with the WAM-5500 autorefractor, respectively.