Objective:
Evaluate transcatheter mitral valve replacement (TMVR) valve-in-valve (VIV) outcomes in three different mitral bioprostheses (of comparable measured internal diameters) under stable hemodynamic and surgical conditions by bench, echocardiographic, computerized tomography (CT), and autopsy comparisons pre- and post-valve implantation in a porcine model under matched controlled conditions.
Background:
Impact of surgical bioprosthesis design on TMVR VIV procedures is unknown.
Methods:
Fifteen similar-sized Yorkshire pigs underwent pre-procedural CT screening. Twelve had consistent anatomic features and underwent implantation of mitral bioprostheses. Four valves from each of three manufacturers were implanted in randomized fashion: 27-mm Epic, 27-mm Mosaic, and 25-mm Mitris, followed by TMVR VIV with 26 Edwards Sapien3. Post-VIV, suprasternal TEE studies were performed to assess hemodynamic function, followed by a gated contrast CT. After euthanasia, animals underwent necropsy for anatomic evaluation.
Results:
All 12 animals had successful VIV implantation with no study deaths. The post vivMitris (3.77 ± 0.36)/(2.2 ± 0.25 mmHg) had the lowest peak/mean trans-mitral gradient and the vivEpic the highest (15.5 ± 2.55)/(7.09 ± 1.13 mmHg). All THVs (transcatheter heart valves) had greatest deformation within the center of the THV frame; with the smallest waist opening area in the vivEpic (329 ± 35.8 mm2 ) and greatest in the vivMitris (414 ± 33.12 mm2 ). Bioprosthetic frames without obvious radiopaque markers resulted in the most ventricular implantation of the THV's anteroseptal frame (Epic: -4.52 ± 0.76 mm), versus the most radiopaque bioprosthesis (Mitris: -1.18 ± 2.95 mm), and higher peak LVOT gradients (Epic: 4.82 ± 1.61 mmHg; Mitris: 2.91 ± 1.47 mmHg).
Conclusions:
The current study demonstrates marked variations in hemodynamics, THV opening area, and anatomic dimensions among measured similarly sized mitral bioprostheses. These data suggest a critical need for understanding the potential impact of variations in bioprosthesis design on TMVR VIV clinical outcomes.
More Related Videos
Related Concept Videos
Mitral Valve Prolapse II: Assessment and Management
Mitral Valve Prolapse I: Introduction
Mitral Stenosis III: Medical Management
Mitral Stenosis II: Clinical features and Diagnostic Tests
Mitral Regurgitation II: Clinical Features and Diagnostic Tests
Mitral Regurgitation I: Introduction