Objective:
To assess the efficacy and safety of closed-loop insulin delivery compared with sensor-augmented pump therapy among older adults with type 1 diabetes.
Research Design And Methods:
This open-label, randomized (1:1), crossover trial compared 4 months of closed-loop versus sensor-augmented pump therapy. Eligible adults were aged ≥60 years, with type 1 diabetes (duration ≥10 years), using an insulin pump. The primary outcome was continuous glucose monitoring (CGM) time in range (TIR; 3.9-10.0 mmol/L).
Results:
There were 30 participants (mean age 67 [SD 5] years), with median type 1 diabetes duration of 38 years (interquartile range [IQR] 20-47), randomized (n = 15 to each sequence); all completed the trial. The mean TIR was 75.2% (SD 6.3) during the closed-loop stage and 69.0% (9.1) during the sensor-augmented pump stage (difference of 6.2 percentage points [95% CI 4.4 to 8.0]; P < 0.0001). All prespecified CGM metrics favored closed loop over the sensor-augmented pump; benefits were greatest overnight. Closed loop reduced CGM time <3.9 mmol/L during 24 h/day by 0.5 percentage points (95% CI 0.3 to 1.1; P = 0.0005) and overnight by 0.8 percentage points (0.4 to 1.1; P < 0.0001) compared with sensor-augmented pump. There was no significant difference in HbA1c between closed-loop versus sensor-augmented pump stages (7.3% [IQR, 7.1-7.5] (56 mmol/mol [54-59]) vs. 7.5% [7.1-7.9] (59 mmol/mol [54-62]), respectively; P = 0.13). Three severe hypoglycemia events occurred during the closed-loop stage and two occurred during the sensor-augmented pump stage; no hypoglycemic events required hospitalization. One episode of diabetic ketoacidosis occurred during the sensor-augmented pump stage; no serious adverse events occurred during the closed-loop stage.
Conclusions:
Closed-loop therapy is an effective treatment option for older adults with long-duration type 1 diabetes, and no safety issues were identified. These older adults had higher TIR accompanied by less time below range during closed loop than during sensor-augmented pump therapy. Of particular clinical importance, closed loop reduced the time spent in hypoglycemic range overnight.
More Related Videos
06:13Author Spotlight: Exploring the Impact of Reduced Resistance Exercise Volume on Metabolic Health
Published on: December 1, 2023
08:32Studying the Hypothalamic Insulin Signal to Peripheral Glucose Intolerance with a Continuous Drug Infusion System into the Mouse Brain
Published on: January 4, 2018
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Crossover Experiments
Crossover designs are performed even with smaller sample sizes since the samples can act as their controls. These are better than simple randomized trials since patients are exposed to all the treatments.
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Insulin: The Receptor and Signaling Pathways