Isokinetic Robotic Device to Improve Test-Retest and Inter-Rater Reliability for Stretch Reflex Measurements in Stroke Patients with Spasticity
Published on: June 12, 2019
Development and reliability of a new system for bedside evaluation of non-volitional knee extension force
Marco Aurélio Vaz1, Matias Fröhlich1, Danton Pereira da Silva Júnior2
1School of Physical Education, Physical Therapy and Dance, Federal University of Rio Grande do Sul, Porto Alegre, RS, Brazil.
Purpose:
Neuromuscular electrical stimulation (NMES) is a widely-used technique for diagnostic and therapeutic purposes. Here we developed and tested the reliability of a new NMES-dynamometer system for bedside evaluation of knee extensor muscle function.
Materials And Methods:
Thirty-two healthy participants (16 men, 16 women; 27±5 years) completed two testing sessions, 7 days apart. On day 1, a single experienced rater, who repeated the evaluation on day 2 with two other raters, completed a standardized testing procedure. Participants were placed supine, with knees flexed and legs connected to the dynamometer. Maximal voluntary knee extensor isometric force (MVF) and supramaximal twitch force (TwF) were obtained.
Results:
High intra-rater intraclass correlation coefficients were observed for both MVF (0.91) and TwF (0.94). MVF and TwF standard error of measurements (8.2%, 5.9%) and minimal detectable changes (16%, 11.6%) were low compared to mean values. High intraclass correlation coefficients were also observed for inter-rater comparisons of MVF (0.89) and TwF (0.86). Standard errors of measurements (MVF: 8.7%, TwF: 5.5%) and minimal detectable changes (MVF: 17.2%, TwF: 10.8%) were similar to intra-rater comparisons.
Conclusion:
The good reliability of the novel NMES-dynamometer system suggests it as an appropriate tool for the bedside evaluation of knee extensor muscle function.