Related Experiment Video
Updated: Oct 11, 2025

Automated Delivery of Microfabricated Targets for Intense Laser Irradiation Experiments
Published on: January 28, 2021
Design and commissioning of a neutron counter adapted to high-intensity laser matter interactions
V Lelasseux1, P-A Söderström2, S Aogaki2
1LULI-CNRS, CEA, UPMC Univ. Paris 06: Sorbonne Université, Ecole Polytechnique, Institut Polytechnique de Paris, F-91128 Palaiseau Cedex, France.
Abstract:
The advent of multi-PW laser facilities world-wide opens new opportunities for nuclear physics. With this perspective, we developed a neutron counter taking into account the specifics of a high-intensity laser environment. Using GEANT4 simulations and prototype testings, we report on the design of a modular neutron counter based on boron-10 enriched scintillators and a high-density polyethylene moderator. This detector has been calibrated using a plutonium-beryllium neutron source and commissioned during an actual neutron-producing laser experiment at the LULI2000 facility (France). An overall efficiency of 4.37(59)% has been demonstrated during calibration with a recovery time of a few hundred microseconds after laser-plasma interaction.