Related Experiment Video
Updated: Oct 11, 2025

06:48
Fluid-cell Raman Spectroscopy for operando Studies of Reaction and Transport Phenomena during Silicate Glass Corrosion
Published on: May 9, 2025
698
A multi-purpose high-pressure and high temperature gas-flow cell for operando optical Raman spectroscopy
Changchun Yang1, Weifeng Huang1, Hang Wei2
1School of Electronic Engineering, Jiujiang University, Jiujiang, Jiangxi Province 332005, China.
The Review of Scientific Instruments
|December 2, 2021
Summary
A versatile in operando optical cell enables Raman spectroscopy under diverse temperature and gas conditions. This advanced setup facilitates in-depth analysis of material reactions for various scientific applications.
Area of Science:
- Materials Science
- Spectroscopy
- Chemical Engineering
Background:
- In operando studies require specialized equipment for real-time analysis.
- Controlling temperature and gas pressure is crucial for understanding reaction mechanisms.
Purpose of the Study:
- To describe a novel multi-purpose in operando optical cell.
- To enable in operando Raman spectroscopy under controlled temperature and gas pressure.
Main Methods:
- The optical cell features precise temperature control (room temperature to 1000°C).
- Gas pressure control ranges from 10⁻¹ Pa to 6 MPa, accommodating various gases (oxidizing and reducing).
- A uniform and stable heating stage ensures reliable experimental conditions.
Main Results:
- The device was successfully calibrated for in operando measurements.
- Raman spectra were collected for different reaction systems.
- The cell demonstrated compatibility with diverse operating temperatures and pressures.
Conclusions:
- The developed optical cell is a valuable tool for in operando Raman spectroscopy.
- It provides a versatile platform for studying chemical reactions under varied conditions.