Patient Directed Recording of a Bipolar Three-Lead Electrocardiogram using a Smartwatch with ECG Function
Published on: December 11, 2019
Effect of the recording condition on the quality of a single-lead electrocardiogram
Satomi Hamada1, Kanae Sasaki2, Hotaka Kito2
1Department of Clinical Laboratory, Tokyo Medical and Dental University (TMDU) Hospital, Tokyo, Japan.
Abstract:
Although many wearable single-lead electrocardiogram (ECG) monitoring devices have been developed, information regarding their ECG quality is limited. This study aimed to evaluate the quality of single-lead ECG in healthy subjects under various conditions (body positions and motions) and in patients with arrhythmias, to estimate requirements for automatic analysis, and to identify a way to improve ECG quality by changing the type and placement of electrodes. A single-lead ECG transmitter was placed on the sternum with a pair of electrodes, and ECG was simultaneously recorded with a conventional Holter ECG in 12 healthy subjects under various conditions and 35 patients with arrhythmias. Subjects with arrhythmias were divided into sinus rhythm (SR) and atrial fibrillation (AF) groups. ECG quality was assessed by calculating the sensitivity and positive predictive value (PPV) of the visual detection of QRS complexes (vQRS), automatic detection of QRS complexes (aQRS), and visual detection of P waves (vP). Accuracy was defined as a 100% sensitivity and PPV. We also measured the amplitude of the baseline, P wave, and QRS complex, and calculated the signal-to-noise ratio (SNR). We then focused on aQRS and estimated thresholds to obtain an accurate aQRS in more than 95% of the data. Finally, we sought to improve ECG quality by changing electrode placement using offset-type electrodes in 10 healthy subjects. The single-lead ECG provided 100% accuracy for vQRS, 87% for aQRS, and 74% for vP in healthy subjects under various conditions. Failure for accurate detection occurred in several motions in which the baseline amplitude was increased or in subjects with low QRS or P amplitude, resulting in low SNR. The single-lead ECG provided 97% accuracy for vQRS, 80% for aQRS in patients with arrhythmias, and 95% accuracy for vP in the SR group. The AF group showed higher baseline amplitude than the SR group (0.08 mV vs. 0.02 mV, P < 0.01) but no significant difference in accuracy for aQRS (79% vs. 81%, P = 1.00). The thresholds to obtain an accurate aQRS were a QRS amplitude > 0.42 mV and a baseline amplitude < 0.20 mV. The QRS amplitude was significantly influenced by electrode placement and body position (P < 0.01 for both, two-way analysis of variance), and the maximum reduction by changing body position was estimated as 30% compared to the sitting posture. The QRS amplitude significantly increased when the inter-electrode distance was extended vertically (1.51 mV for vertical extension vs. 0.93 mV for control, P < 0.01). The single-lead ECG provided at least 97% accuracy for vQRS, 80% for aQRS, and 74% for vP. To obtain stable aQRS in any body positions, a QRS amplitude > 0.60 mV and a baseline amplitude < 0.20 mV were required in the sitting posture considering the reduction induced by changing body position. Vertical extension of the inter-electrode distance increased the QRS amplitude.
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Electrophysiology of Normal Cardiac Rhythm
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...