Related Experiment Video
Updated: Oct 11, 2025

X-ray Dose Reduction through Adaptive Exposure in Fluoroscopic Imaging
Published on: September 11, 2011
Revisiting dynamic range and image enhancement ability of contemporary digital radiographic systems
Luiz Eduardo Marinho-Vieira1, Luciano Augusto Cano Martins1, Deborah Queiroz Freitas1
1Division of Oral Radiology, Department of Oral Diagnosis, Piracicaba Dental School, University of Campinas (UNICAMP), Piracicaba-SP, Brazil.
Objectives:
To assess the dynamic range and enhancement ability of radiographs acquired with contemporary digital systems.
Methods:
Five repeated periapical radiographs of human mandibles with an aluminium step-wedge were acquired using two sensor-based and three photostimulable phosphor plate-based systems and an X-ray unit at ten exposure times 0.020, 0.032, 0.063, 0.080, 0.100, 0.200, 0.320, 0.400, 0.500, and 0.630 s. All images had their brightness and contrast enhanced by two experienced oral and maxillofacial radiologists in consensus and were exported as both the original and enhanced file formats. Mean grey values were obtained from the aluminium steps and tabulated with their corresponding thicknesses for each exposure time, digital radiographic system, and file format. Images with saturated steps were excluded and the mean grey values from the remaining images were averaged to assess image brightness and the angular coefficient of the linear trendlines was generated from the relationship between mean grey values and their corresponding aluminium thicknesses to assess image contrast. Brightness and contrast values were compared using two-way ANOVA with post-hoc Tukey (α = 0.05).
Results:
Photostimulable phosphor plate-based digital radiographic systems had a broader dynamic range. Longer exposure times produced original images with lower brightness and variable contrast (p < 0.05). Subjective enhancement significantly increased or reduced brightness and/or contrast in some systems (p < 0.05).
Conclusions:
Contemporary digital radiographic systems present different dynamic ranges and exposure-related brightness and contrast. Image enhancement may be a valuable tool at slightly suboptimal exposure times.
Related Concept Videos
X-ray Imaging
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...