Deep learning-based robust automatic non-invasive measurement of blood pressure using Korotkoff sounds
Ji-Ho Chang1, Il Doh2,3
1Division of Physical Metrology, Korea Research Institute of Standards and Science (KRISS), Daejeon, Republic of Korea.
Abstract:
This paper proposes a method that automatically measures non-invasive blood pressure (BP) based on an auscultatory approach using Korotkoff sounds (K-sounds). There have been methods utilizing K-sounds that were more accurate in general than those using cuff pressure signals only under well-controlled environments, but most were vulnerable to the measurement conditions and to external noise because blood pressure is simply determined based on threshold values in the sound signal. The proposed method enables robust and precise BP measurements by evaluating the probability that each sound pulse is an audible K-sound based on a deep learning using a convolutional neural network (CNN). Instead of classifying sound pulses into two categories, audible K-sounds and others, the proposed CNN model outputs probability values. These values in a Korotkoff cycle are arranged in time order, and the blood pressure is determined. The proposed method was tested with a dataset acquired in practice that occasionally contains considerable noise, which can degrade the performance of the threshold-based methods. The results demonstrate that the proposed method outperforms a previously reported CNN-based classification method using K-sounds. With larger amounts of various types of data, the proposed method can potentially achieve more precise and robust results.
Related Concept Videos
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Korotkoff Sounds
During blood pressure assessment, inflating the cuff 30 millimeters of mercury above the patient's systolic blood pressure...
Assessment of blood pressure in brachial artery(two-step method)
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: