Simon N van Laarhoven1, Abraham H J van Eerden1, Gijs G van Hellemondt1
1Department of Orthopedic Surgery, Sint Maartenskliniek, Nijmegen, the Netherlands.
Background:
Clinical observations revealed higher rates of aseptic loosening for hybrid fixated rotating hinge knee implants compared to fully cemented ones. We hypothesize that the use of a fully cemented fixation technique had a higher survival rate for aseptic loosening compared to a hybrid fixation technique in a rotating hinge knee implant.
Methods:
All procedures of patients who were treated with the RT-PLUS rotating hinge knee implant (Smith & Nephew, Memphis, TN) between 2010 and 2018 were included. Primary outcome was revision for aseptic loosening. Kaplan-Meier survivorship and Cox proportional hazard regression analysis were performed to calculate survival rates and hazard ratios.
Results:
A total of 275 hinge knee implants were placed in 269 patients (60 primary procedures, 215 revisions). Median follow-up was 7.3 ± 3.9 years. In total, 24 components (16 hybrid femur, 2 fully cemented femur, 6 hybrid tibia; all revision procedures) in 19 patients were revised for aseptic loosening. Kaplan-Meier survivorship analysis showed superior survival rates of fully cemented components (femur 97.1%; tibia 100%) compared to hybrid fixated components (femur 89.5%; tibia 95.9%) at the 10-year follow-up. Multivariate Cox hazard analysis showed a significantly higher risk of aseptic loosening for hybrid fixated components, a prior stemmed component and the femoral component.
Conclusion:
Fully cemented fixation showed superior survival rates for aseptic loosening compared to hybrid fixation in a single design rotating hinge knee implant. A prior stemmed component appears to be a risk factor for aseptic loosening and the femoral component seems to be more prone to loosening.