Radiographic and endoscopic sensitivity in detecting lower esophageal mucosal ring
AJR. American Journal of Roentgenology
|August 1, 1986
Summary
Radiography is more accurate than endoscopy for detecting lower esophageal mucosal rings. Barium esophagram identified 95% of rings, while endoscopy detected only 58%, especially missing larger ones.
Area of Science:
- Gastroenterology
- Radiology
- Medical Imaging
Background:
- Lower esophageal mucosal ring is a condition affecting the esophagus.
- Accurate detection is crucial for diagnosis and management of dysphagia.
Purpose of the Study:
- To compare the diagnostic sensitivity of radiography and endoscopy for lower esophageal mucosal rings.
- To evaluate factors influencing detection rates for each modality.
Main Methods:
- Retrospective analysis of 60 patients with lower esophageal mucosal rings.
- Comparison of barium esophagram (prone full-column and double-contrast techniques) with endoscopy.
- Correlation of detection rates with ring caliber and endoscope size.
Main Results:
- Barium esophagram detected 95% of rings; double-contrast technique detected 46%.
- Endoscopy detected 58% of rings, with sensitivity decreasing for larger rings (≤13mm: 82%, 14-19mm: 54%, ≥20mm: 25%).
- Dysphagia was more common in patients with endoscopically undetected rings; larger endoscopes improved detection rates.
Conclusions:
- Radiography, particularly barium esophagram, is more sensitive than endoscopy for detecting lower esophageal mucosal rings.
- Initial radiographic examination is recommended for patients with dysphagia and suspected esophageal narrowing.