Ablation with zero-fluoroscopy of premature ventricular complexes from aortic sinus cusps: A single-center experience

Pablo J Sánchez-Millán1,2, Guillermo Gutiérrez-Ballesteros1,2, Manuel Molina-Lerma1,2

  • 1Arrhythmia Unit Hospital Universitario Virgen de las Nieves Granada Spain.

Journal of Arrhythmia
|December 10, 2021
PubMed

Insights

Zero-fluoroscopy (ZF) catheter ablation for aortic sinus cusp premature ventricular complexes (ASC-PVC) using 3D intracardiac echocardiography integrated electroanatomical mapping is feasible, effective, and safe, avoiding traditional coronary catheterization.

Area of Science:

  • Cardiology
  • Electrophysiology
  • Medical Imaging

Background:

  • Catheter ablation for premature ventricular complexes originating from the aortic sinus cusps (ASC-PVC) typically requires coronary catheterization.
  • Limited data exists on zero-fluoroscopy (ZF) techniques for ASC-PVC mapping and ablation.
  • Conventional methods involve localizing the coronary artery ostium (CAO) via coronary catheterization.

Purpose of the Study:

  • To evaluate the efficacy and safety of a ZF approach for ASC-PVC ablation.
  • To assess the utility of 3D intracardiac echocardiography integrated into an electroanatomical mapping system (ICE 3D-EAM) for guidance.

Main Methods:

  • An observational study comparing a ZF group (n=10) with a control group (n=11) using conventional methods.
  • Patients with ASC-PVC underwent ablation guided by ICE 3D-EAM in the ZF group.
  • Clinical, efficacy, and safety outcomes were assessed acutely and at 3-month follow-up.

Main Results:

  • No fluoroscopy was needed in the ZF group.
  • Acute success rates were 80% for ZF versus 55% for the control group (P=.36).
  • Recurrence rates were 30% in the ZF group and 27% in the control group (P=1). One non-severe complication occurred in the ZF group.

Conclusions:

  • ZF catheter ablation of ASC-PVC guided by ICE 3D-EAM is a feasible and safe alternative.
  • This approach demonstrates effectiveness in treating ASC-PVC without fluoroscopy.
Abstract