Related Experiment Video
Updated: Oct 10, 2025

Normothermic Negative Pressure Ventilation Ex Situ Lung Perfusion: Evaluation of Lung Function and Metabolism
Published on: February 14, 2022
Performance of the Masi Peruvian ventilator at high altitude
Abstract:
In response to Covid-19 crisis, 310 Masi ventilators were produced and validated in Lima, Peru, according to applicable standards. Four of them, were transported to Puno, in order to strengthen ICU Services there, but this set a major challenge to Masi team as effects of altitude on ventilators were unknown. Once there, ventilators were acclimated and calibrated. Volume tidal, I:E ratio, respiratory frequency and PEEP were tested, all of them presenting errors under 15%, except for tidal volume, for which a 25% negative correction was applied. After the installation of a new version of Masi software, parameters were tested again, all of them presenting results with errors below 15%, which allowed the Masi team to take them to ICU services for use.Clinical Relevance- Masi Peruvian Ventilators are able to perform according to their specifications at extremely high altitude, after the adequate calibration. These devices are an alternative to treat COVID-19 patients in the middle of the crisis.
Related Concept Videos
Mechanical Ventilation I: Indication and Settings
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Respiratory Capacities
One key metric is the Inspiratory Capacity (IC), which represents the maximum amount of air that can be inhaled with full effort. IC is calculated by summing the tidal volume and inspiratory reserve volume, typically ranging from 2.4 to 3.6 liters.