Personalization of pulse arrival time based blood pressure surrogates through single spot check measurements
Objective:
We investigate the effect of selective single parameter personalization on the performance of multi-parameter models for pulse arrival time (PAT) based blood pressure (BP) surrogates.
Methods:
Our data set stems from 15 surgery patients, and we selected from each patient 5 segments of 30 min length each. We evaluate the root mean squared BP tracking error of the two models with and without single parameter personalization. We further compare the BP tracking performance to a surrogate-free sample-and-hold approach, e.g., as afforded by conventional non-invasive blood pressure (NIBP) oscillometry.
Results:
Parameter personalization is key to realizing a tracking performance benefit of PAT-based BP surrogates. The highest tracking error reduction of about 3.7 mmHg with respect to a sample-and-hold approach was reached with a personalized model which is linear in the pulse wave velocity domain. It achieves an estimation error of 7.8 mmHg with respect to a continuously measured invasive reference.Clinical Relevance-We give a performance analysis of PAT-based BP surrogates which are personalized to a patient with a single NIBP spot measurement. We show for surgery patients that patient-specific personalization enables continuous beat-to-beat BP monitoring over 30 min intervals with a average root mean squared error of less than 8 mmHg.
Related Concept Videos
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Errors occurring during blood pressure monitoring
Several factors...
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)
Measurement of Blood Pressure