An investigation of the individualized, two-point calibration method for cuffless blood pressure estimation using
Background:
The use of wearable cuffless blood pressure (BP) devices is becoming commercially prevalent with little published validation information. Most devices rely, at least in part, on the relationship between pulse arrival time (PAT) and BP, a theoretical fundamental relationship that was first commercially exploited in 1993 with the release of the Casio BP-100 digital watch.
Objective:
This study explored the PAT method of BP estimation in a commercial device where it first began, the Casio BP-100 (Model No. 900) digital watch, which employs an individualized, two-point calibration method. Device accuracy was determined by comparison to a conventional cuff-based BP device measurements.
Methods:
Twenty participants (11 female, 9 male) had BP measured using both devices at rest, during a 5-minute isometric hand-grip exercise and at 1-minute post-exercise.
Results:
Due to bidirectional scatter of BP estimation by the BP-100 device, there was no significant difference between the reference device and the BP-100. The devices showed poor correlation for both systolic BP (SBP) (R=0.36, p=0.13) and diastolic BP (DBP) (R=0.044, p=0.37). However, on average the watch was able to provide correct directional changes in SBP but not DBP with exercise.
Conclusions:
Despite being an industry first, the Casio BP-100 watch employed a method that gives a great chance of accuracy: a two point, individualized calibration method - more detailed than calibration methods in more modern devices. The watch, on average across a cohort, provided some information on BP directional change but was uncorrelated with cuff-based BP measurement. If the utility of beat-by-beat BP estimation is to be utilized, limitations of this method need to be addressed.
Related Concept Videos
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure