Application of chromium catalysis technology to compound-specific hydrogen isotope analysis of natural gas samples
Lan-Tian Xing1, Zhong-Ping Li1, Li Xu2
1Research Center for Oil and Gas Resources, Northwest Institute of Eco-Environment and Resources, Chinese Academy of Sciences, Lanzhou 730000, China.
Abstract:
An easy-to-use and efficient approach for the catalytic pyrolysis of hydrocarbon gas here is proposed and tested, which allowed the precise measurement of δ2H. This method requires the reaction tube filled with twisted chromium wire, establishing a gas chromatography-Chromium wire/high temperature conversion-isotope ratio mass spectrometer (GC‒Cr/HTC‒IRMS) system. The approach described here has many remarkable advantages over other techniques. In comparison to traditional carbon-coating pyrolysis technique, this proposed method greatly reduces the pyrolysis temperature of the hydrocarbon gas, especially methane, with temperatures reduced from 1450 to 1250 °C, while still ensuring excellent analytical precision. Therefore, hydrogen isotope analysis of natural gas can be carried out at a lower cracking temperature for a longer time, with the advantage of increasing the service life of the HTC furnace. Meanwhile, no pretreatment or activation is required while putting the Cr/HTC system to use. Further, the accuracy of δ2H values obtained using this new method is better than using the traditional method. Our experiments demonstrate that the test errors of this novel method and the conventional method are within ±5.5‰ and ±6.5‰, respectively.
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...