Wide Variability in the Sensitivity and Specificity of Rotavirus Immunoassay Diagnostic Kits in Practice
Rouba Shaker1, Ebla Abdalrahman1, Zainab Ali1
1Department of Pediatrics and Adolescent Medicine, American University of Beirut, Faculty of Medicine, Beirut, Lebanon.
Introduction:
Most hospitals rely on rapid antigen-detection kits for the diagnosis of rotavirus infection. Several small studies reviewed the sensitivity and specificity of some of these kits. These studies showed discrepancy in results obtained for sensitivity and specificity that varied according to the type of kit used, area of study, and type of test used as standard for diagnosis of rotavirus infection. The objective of the study is to determine the sensitivity and specificity of five commonly used rotavirus immunoassay kits in comparison to RT-PCR as standard.
Methodology:
Stool samples (N = 1,414) collected from children under 5 years of age hospitalized with gastroenteritis were tested for rotavirus by immunoassay kits and RT-PCR in a prospective hospital-based surveillance study conducted at 7 centers in Lebanon. Concordance and discrepancy between the two methods was used to calculate sensitivity and specificity, using RT-PCR as the "gold standard".
Results:
The sensitivity and specificity were respectively 95.08% and 86.62% for the SD Bioline® (Standard Diagnostics, Inc, South Korea) kit calculated on 645 samples, 65.86% and 45.90% for the VIROTECT® (Trinity Biotech, Ireland) kit calculated on 327 samples, 83.9% and 64.2% for the Rota-Strip (C-1001) (Coris Bioconcept, Belgium) calculated on 95 samples, 52.3% and 10.9% for the Acon® (Acon Laboratories, Inc, California, USA) kit calculated on 122 samples, 68.1% and 20% for the VIKIA® Rota-Adéno (Biomerieux, France) kit calculated on 32 samples.
Conclusion:
A wide discrepancy was detected between the calculated and advertised sensitivity and specificity for most of the kits.