A LC-MS/MS method with electrospray ionization and atmospheric pressure chemical ionization source for analysis of
Avinash Dalmia1, Erasmus Cudjoe2, Jacob Jalali3
1Perkin Elmer, Inc., 710 Bridgeport Ave., Shelton, CT, 06484, USA. Avinash.dalmia@perkinelmer.com.
Background:
Pesticide testing for hemp has traditionally focused on techniques like QuEChERS with dSPE and SPE which demand time-consuming sample preparation, typically resulting in poor recovery rates for some pesticides, and requires the use of both LC-MS/MS and GC-MS/MS based instruments to cover the analysis for all regulated pesticides. In this study, we describe a streamlined approach for working with LC-MS/MS featuring a dual electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI) sources using solvent extraction for faster and easier sample preparation and with 80-120% recovery for the analysis of all of 66 pesticides (regulated by California state in cannabis) with low detection limits in hemp.
Methods:
A simple solvent extraction with acetonitrile was used to extract pesticides from hemp. A LC-MS/MS system with dual ESI and APCI source was used to determine sensitivity for the analysis of 66 pesticides in hemp matrix, 62 pesticides were analyzed using an 18-min LC-MS/MS method with an ESI source and the other 4 pesticides were measured using a 6-min LC-MS/MS method with an APCI source.
Results:
The limit of quantitation (LOQ) of all 66 pesticides in hemp was in the range of 0.0025-0.1 μg/g which was well below the California state action limits of these analytes in cannabis products. A simple, fast, and cost-effective solvent extraction method was used for sample preparation to get good recovery in the range of 80-120% with RSD less than 20%. The unique ionization mechanism of chlorinated pesticides such as pentachloronitrobenzene using the LC-MS/MS system with an APCI source was elucidated. The proficiency test report generated with the LC-MS/MS method showed acceptable results for all of 66 pesticides in hemp with all of th z scores less than 2 with no false positives and negatives. The stability data collected over 5 days showed RSD less than 20% for 66 pesticides in hemp, and this demonstrated the robustness of the LC-MS/MS system used in this work.
Conclusions:
A LC-MS/MS method with dual ESI and APCI sources was developed for the analysis of 66 pesticides in hemp. The recovery of all pesticides from a hemp matrix was in the acceptable range of 80-120% with RSD less than 20%.
More Related Videos
04:17Analysis of Organochlorine Pesticides in a Soil Sample by a Modified QuEChERS Approach Using Ammonium Formate
Published on: January 20, 2023
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Chemical Ionization (CI) Mass Spectrometry
High-Performance Liquid Chromatography: Types of Detectors
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...