Related Experiment Video
Updated: Oct 10, 2025

CMAP Scan MUNE MScan - A Novel Motor Unit Number Estimation MUNE Method
Published on: June 7, 2018
Assessing inter-rater reproducibility in MScanFit MUNE in a 6-subject, 12-rater "Round Robin" setup
Daniel Mosgaard Sørensen1, Hugh Bostock2, Martin Ballegaard3
1Department of Clinical Neurophysiology, Aarhus University Hospital, Denmark.
Objective:
To assess the inter-rater reliability of MScanFit MUNE using a "Round Robin" research design.
Methods:
Twelve raters from different centres examined six healthy study participants over two days. Median, ulnar and common peroneal nerves were stimulated, and compound muscle action potential (CMAP)-scans were recorded from abductor pollicis brevis (APB), abductor digiti minimi (ADM) and anterior tibial (TA) muscles respectively. From this we calculated the Motor Unit Number Estimation (MUNE) and "A50", a motor unit size parameter. As statistical analysis we used the measures Limits of Agreement (LOA) and Coefficient of Variation (COV). Study participants scored their perception of pain from the examinations on a rating scale from 0 (no pain) to 10 (unbearable pain).
Results:
Before this study, 41.6% of the raters had performed MScanFit less than five times. The mean MUNE-values were: 99.6 (APB), 131.4 (ADM) and 126.2 (TA), with LOA: 19.5 (APB), 29.8 (ADM) and 20.7 (TA), and COV: 13.4 (APB), 6.3 (ADM) and 5.6 (TA). MUNE-values correlated to CMAP max amplitudes (R2-values were: 0.463 (APB) (p<0.001), 0.421 (ADM) (p<0.001) and 0.645 (TA) (p<0.001)). The average perception of pain was 4.
Discussion:
MScanFit indicates a high level of inter-rater reliability, even with only limited rater experience and is overall reasonably well tolerated by patients. These results may indicate MScanFit as a reliable MUNE method with potential as a biomarker in drug trials.
More Related Videos
08:52Software-Assisted Quantitative Measurement of Osteoarthritic Subchondral Bone Thickness
Published on: March 18, 2022
12:50Lesion Explorer: A Video-guided, Standardized Protocol for Accurate and Reliable MRI-derived Volumetrics in Alzheimer's Disease and Normal Elderly
Published on: April 14, 2014