Accuracy of ultra-high-resolution computed tomography with a 0.3-mm detector for quantitative assessment of coronary
Minoru Yamada1, Yoshitake Yamada1, Takehiro Nakahara1
1Department of Radiology, Keio University School of Medicine, 35 Shinanomachi, Shinjuku-ku, Tokyo, 160-8582, Japan.
Background:
The development of ultra-high-resolution CT (U-HRCT) is expected to improve the accuracy of coronary stenosis evaluation. This study aimed to evaluate the accuracy of the stenosis severities of coronary artery phantoms estimated using U-HRCT by comparing them to those estimated with conventional CT.
Methods:
Coronary artery phantoms with non-calcified and calcified lesions were scanned with conventional CT (64-row × 0.625 mm) and U-HRCT (32-row × 0.3125 mm). The coronary artery phantoms had lumen diameters of 2.0, 3.0, and 4.0 mm with non-calcified lesions representing 0%, 25%, 50%, and 75% stenosis and 3.0 and 4.0 mm with calcified lesions representing 0%, 25%, 50%, and 75% stenosis. The lumen diameters at the stenotic and non-stenotic regions were measured, and the stenosis severities were compared with the true values.
Results:
For non-calcified lesions, conventional CT significantly underestimated the stenosis severity in the phantom showing 75% stenosis with lumen diameters of 2.0 and 3.0 mm (p < 0.05), while the estimated stenosis severities were not significantly different from the true values at all settings with U-HRCT. For the calcified lesions, conventional CT overestimated the stenosis severities at all settings (p < 0.05), while U-HRCT yielded estimations closer to the true values, although still with some overestimation (p < 0.05).
Conclusion:
By using U-HRCT, the estimated stenosis severities of the coronary artery with non-calcified lesion become almost equal to the true value, while those with calcified lesion are still overestimated although they become closer to the true value.
More Related Videos
09:57Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
06:57Author Spotlight: Advancing Cardiovascular Imaging - Introducing the Spatially Weighted Calcium Score for Early Disease Detection
Published on: September 22, 2023
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Acute Coronary Syndrome III: Diagnostic Studies