Riata lead failure manifests as electrical storm with lead parameters showing unique postural variations: a case
Abhimanyu Uppal1, Sanjeev Kathuria1, Bhushan Shah1
1Department of Cardiology, GB Pant Institute of Post Graduate Education and Research, MAMC, JLN Road, New Delhi 110002, India.
Background:
Riata implantable cardioverter-defibrillator (ICD) leads are prone to a unique type of mechanical lead failure causing conductor externalization (CE) which may be complicated by a delayed-onset electrical lead failure (ELF).
Case Summary:
A 60-year-old male with symptomatic, severe ischaemic cardiomyopathy, and atrial fibrillation following a prior anterior wall myocardial infarction received a dual-chamber ICD with 7F-RiataST ventricular lead as a primary prevention strategy against sudden cardiac death in 2008. In 2017, a pulse generator replacement was performed for elective replacement indicator status. At that time, CE was noted in the ventricular lead but the electrical lead parameters were normal, hence lead replacement was decided against and the patient was closely followed up thereafter. Four years later, the patient presented with multiple ICD shocks within 48 h. Implantable cardioverter-defibrillator interrogation showed noise on the ventricular electrogram (EGM) channel that was detected as ventricular fibrillation (VF) episodes, triggering inappropriate ICD therapy (five ICD detected VF events within 24 h triggering three antitachycardia pacing therapies and one shock). Lead impedance and R-wave amplitude were within normal range in supine position but dramatically worsened in sitting posture. A new ventricular lead was implanted and the old lead abandoned. The patient has not experienced any device therapy in the follow-up period.
Discussion:
An electrically inert CE of Riata ICD leads needs close follow-up because an ELF may occur even after several years. A careful analysis of EGMs including postural changes in lead parameters can aid in detection and better characterization of underlying electrical dysfunction following CE.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
09:16Use of a Wireless Video-EEG System to Monitor Epileptiform Discharges Following Lateral Fluid-Percussion Induced Traumatic Brain Injury
Published on: June 21, 2019
Related Concept Videos
Dysrhythmias V: Evaluating Dysrhythmias
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
Dysrhythmias II: Classification of Tachyarrhythmias
Dysrhythmias III: Characteristics of Dysrhythmias