Related Experiment Video
Updated: Oct 10, 2025

Potentiodynamic Corrosion Testing
Published on: September 4, 2016
Adverse Tissue Reactions and Metal Ion Behavior After Small-Head Metasul Hip Arthroplasty: A Long-Term Follow-Up
Tsunehito Ishida1, Toshiyuki Tateiwa1, Yasuhito Takahashi1,2
1Department of Orthopedic Surgery, Tokyo Medical University, Tokyo, Japan.
Objective:
To investigate the long-term survivorship, incidence of adverse reactions to metal debris (ARMD), and metal ion behavior in patients who underwent small-head Metasul metal-on-metal (MoM) total hip arthroplasty (THA).
Methods:
Between February 1998 and September 2003, a retrospective study was performed on 43 consecutive patients (43 hips) who underwent unilateral cementless Metasul MoM THAs at our institution. Of them, 35 patients (nine males and 26 females) who were available for follow-up more than 15 years after THA were enrolled in this study and underwent metal artifact reduction sequence magnetic resonance imaging (MARS-MRI) to identify ARMD. The mean age at surgery of the patients was 59.7 years old (range, 31-83). Clinical and radiographic outcomes were evaluated retrospectively. Clinical examinations were conducted using the Harris Hip Score (HHS). Serum cobalt (Co) and chromium (Cr) ion levels and Co/Cr ratio were assessed at different postoperative periods of <5, 5-10, 11-14, and ≥15 years.
Results:
The mean follow-up period for the 35 patients included was 18.1 years (range, 15-22). The mean HHS significantly improved from 44.6 ± 11.3 points preoperatively to 89.4 ± 7.9 points at the final follow-up (P < 0.0001). ARMD was found in 20% of the patients using MARS-MRI. No signs of stem loosening were found clinically or radiographically, whereas cup loosening and ARMD were observed in three patients (9%), for whom revision THAs were performed. The Kaplan-Meier survival rates with revision for any reason as the endpoint were 90.9% at 5 years, 84.8% at 10 years, 84.8% at 15 years (95% CI, 67.1-93.6), and 70.3% at 20 years (95% CI, 43.6-87.0). The survival rates with revision for ARMD as the endpoint were 100% at 5 years, 96.6% at 10 years, 96.6% at 15 years (95% CI, 77.2-99.7), and 80.1% at 20 years (95% CI, 45.3-95.2). Serum Co ion level peaked at 5-10 years after THA, which was significantly higher than that <5 years; however, it decreased to the initial level after 15 years. In contrast, serum Cr ion level significantly increased at 5-10 years and then remained almost constant. Significant differences in Cr ion levels (1.0 vs 2.0 μg/L, P = 0.024) and Co/Cr ratio (1.3 vs 0.9, P = 0.037) were found between non-ARMD and ARMD patients at >11 years postoperatively.
Conclusion:
Our results suggest that increased Cr ion levels and decreased Co/Cr ratio may be signs of ARMD in patients who underwent small-head Metasul MoM THA.
Insights
Long-term follow-up of small-head Metasul metal-on-metal total hip arthroplasty shows improved function. Increased chromium ion levels and a decreased cobalt/chromium ratio may indicate adverse reactions to metal debris.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Clinical Toxicology
Background:
- Metal-on-metal (MoM) total hip arthroplasty (THA) has raised concerns regarding adverse reactions to metal debris (ARMD).
- Small-head Metasul MoM THA implants require long-term evaluation for survivorship and safety.
Purpose of the Study:
- To assess the long-term outcomes of small-head Metasul MoM THA.
- To investigate the incidence of ARMD and metal ion behavior.
- To identify potential biomarkers for ARMD.
Main Methods:
- Retrospective study of 35 patients with Metasul MoM THA followed for over 15 years.
- Clinical and radiographic assessments including Harris Hip Score (HHS).
- Metal artifact reduction sequence MRI (MARS-MRI) for ARMD detection.
- Serum cobalt (Co) and chromium (Cr) ion level analysis at various postoperative intervals.
Main Results:
- Mean HHS significantly improved post-THA (44.6 to 89.4, P < 0.0001).
- ARMD detected in 20% of patients; 9% experienced cup loosening and ARMD requiring revision.
- Kaplan-Meier survival rates with revision for any reason were 70.3% at 20 years.
- Serum Cr levels increased post-THA and remained elevated; Co levels peaked at 5-10 years and then decreased.
- Higher Cr levels and lower Co/Cr ratio were associated with ARMD (P < 0.05).
Conclusions:
- Small-head Metasul MoM THA demonstrates good long-term functional improvement.
- Increased serum chromium ion levels and a decreased cobalt/chromium ratio may serve as indicators for ARMD.
- Continued monitoring for ARMD is crucial in patients with MoM THA.