Efficient SARS-CoV-2 Quantitative Reverse Transcriptase PCR Saliva Diagnostic Strategy utilizing Open-Source Pipetting Robots
Published on: February 11, 2022
Performance Accuracy of a Laboratory-Developed Real-Time RT-PCR Method for Detection of SARS-CoV-2 in Self-Collected
D Jane Hata1, Edward L White2, Michaela M Bridgeman2
1Department of Laboratory Medicine and Pathology, Mayo Clinic, Jacksonville, FL, USA Hata.donna@mayo.edu.
Objective:
The ongoing COVID-19 pandemic caused by SARS-CoV-2 has challenged diagnostic laboratories to re-examine traditional methods for collecting specimens and sample types used in molecular testing. Our goal was to demonstrate that saliva can be used for detecting SARS-CoV-2 and correlates well with established molecular methods using nasopharyngeal (NP) swabs.
Methods:
We examined use of a saliva collection device in conjunction with a laboratory-developed real-time reverse transcription-polymerase chain reaction (LDPCR) method for detecting SARS-CoV-2 in a symptomatic population and compared results with 2 US Food and Drug Administration (FDA)-approved methods (emergency use authorization [EUA]) that use specimens from NP swabs.
Results:
The sensitivity of LDPCR compared with the reference methods was 75.0% (21/28); specificity, 98.1% (104/106). When cycle threshold values were compared between paired specimens using the LDPCR and a EUA reverse transcription PCR method, both targeting the open-reading frame gene, the mean value for saliva was 4.66 cycles higher than for NP specimens.
Conclusion:
Use of self-collected saliva in conjunction with an LDPCR for SARS-CoV-2 compared favorably with 2 FDA EUA methods using NP swabs. The use of an alternative sample type and assay method will aid in expanding the availability of testing during the ongoing COVID-19 pandemic.