Related Experiment Video
Updated: Oct 9, 2025

Treatment with Locking Intramedullary Nailing for Intertrochanteric Fracture of the Femur Utilizing a New Awl with a Distal Positioner
Published on: June 6, 2025
The Relationship Between Arthroplasty Surgeons' Experience Level and Optimal Cable Tensioning in the Fixation of
Kerem Başarır1, Mahmut Kalem1, Ercan Şahin2
1Department of Orthopedics and Traumatology, University of Ankara, Ankara University Medicine Faculty, Ankara, Turkey.
Introduction:
In this study, our aim was to examine the relationship between the arthroplasty surgeons' experience level and their aptitude to adjust the cable tension to the value recommended by the manufacturer when asked to provide fixation with cables in artificial bones that underwent extended trochanteric osteotomy (ETO).
Materials And Methods:
A custom-made cable tensioning device with a microvoltmeter was used to measure the tension values in Newtons (N). An ETO was performed on 4 artificial femur bones. Surgeons at various levels of experience attending the IXth National Arthroplasty Congress were asked to fix the osteotomized fragment using 1.7-mm cables and the tensioning device. The participants' demographic and experience data were investigated and recorded. The surgeons with different level of experience repeated the tensioning test 3 times and the average of these measurements were recorded.
Results:
In 19 (35.2%) of the 54 participants, the force applied to the cable was found to be greater than the 490.33 N (50 kg) value recommended by the manufacturer. No statistically significant difference was determined between the surgeon's years of experience, the number of cases, and the number of cables used and the tension applied over the recommended maximum value (P = .475, P = .312, and P = .691, respectively).
Conclusions:
No significant relationship was found between the arthroplasty surgeon's level of experience and the adjustment of the cable with the correct tension level. For this reason, we believe that the use of tensioning devices with calibrated tension gauges by orthopedic surgeons would help in reducing the number of complications that may occur due to the cable.