Related Experiment Video
Updated: Jul 28, 2026

Comprehensive Endovascular and Open Surgical Management of Cerebral Arteriovenous Malformations
Published on: October 20, 2017
Safety and efficacy of the Mynx Control vascular closure device in peripheral arterial procedures: A prospective
Athanasios Diamantopoulos1,2, Romman Nourzaie1, Doug Mulholland1
1Department of Interventional Radiology, 8945Guys' and St. Thomas' NHS Foundation Trust, London, UK.
Objectives:
This article aims to present a prospective study investigating the safety and efficacy of the Mynx Control extravascular closure device (Cordis Corporation, FL, USA), for femoral arterial closure in patients undergoing peripheral arterial procedures.
Methods:
Between January 2020 and February 2021, 100 Mynx Control devices were deployed in 91 consecutive patients (mean age: 67.5 ± 16.9 years) who underwent peripheral arterial procedures. We used ultrasound and/or fluoroscopy during Mynx Control deployment and ultrasound post-procedure to detect complications. Femoral artery punctures included 62 (62%) antegrade and 38 (38%) retrograde punctures. The mean activated clotting time at time of device deployment was 221s. The primary endpoints were technical success, device failure, and complication rates up to 30 days.
Results:
A 5F vascular sheath was used in 43 cases (43%) (36 (36%) 6F and 21 (21%) 7F). The majority of our cases had antegrade access (62%). Overall technical success rate was 97% in both antegrade and retrograde cases. In total, there were 4 minor complications: 3 (3%) cases of pseudoaneurysm and 1 (1%) case of haematoma. No major complications were recorded post-procedure or 30 days post index procedure.
Conclusion:
The Mynx Control vascular closure device is safe and effective in achieving haemostasis in patients undergoing antegrade and retrograde peripheral angioplasty procedures.