Fat-Water Phantoms for Magnetic Resonance Imaging Validation: A Flexible and Scalable Protocol
Published on: September 7, 2018
Proton magnetic resonance spectroscopy thermometry: Impact of separately acquired full water or partially suppressed
Zhengchao Dong1,2, Matthew S Milak1,2, J John Mann1,2,3
1Department of Psychiatry, Columbia University, College of Physicians and Surgeons, New York, New York, USA.
Abstract:
In proton magnetic resonance spectroscopy (1 H MRS) thermometry, separately acquired full water and partially suppressed water are commonly used for measuring temperature. This paper compares these two approaches. Single-voxel 1 H MRS data were collected on a 3-T GE scanner from 26 human subjects. Every subject underwent five continuous MRS sessions, each separated by a 2-min phase. Each MRS session lasted 13 min and consisted of two free induction decays (FIDs) without water suppression (with full water [FW or w]) and 64 FIDs with partial water suppression (with partially suppressed water [PW or w']). Frequency differences between the two FWs, the first two PWs, the second FW and the first PW (FW2 , PW1 ), or between averaged water ( ) and N-acetylaspartate (NAA), were measured. Intrasubject and intersubject variations of the frequency differences were used as a metric for the error in temperature measurement. The intrasubject variations of frequency differences between FW2 and PW1 , calculated from the five MRS sessions for each subject, were larger than those between the two FWs or between the first two PWs (p = 1.54 x 10-4 and p = 1.72 x 10-4 , respectively). The mean values of intrasubject variations of for all subjects were 4.7 and 4.5 times those of and , respectively. The intrasubject variations of the temperatures based on frequency differences, or ( ), were about 2.5 times greater than those based on averaged water and NAA frequencies ). The mean temperature measured from ) (n = 26) was 0.29°C lower than that measured from and was 0.83°C higher than that from ( ). It was concluded that the use of separately acquired unsuppressed or partially suppressed water signals may result in large errors in frequency and, consequently, temperature measurement.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Chemical Shift: Internal References and Solvent Effects
The internal reference compound generally used in NMR spectroscopy is tetramethylsilane (TMS). TMS is preferred because it is chemically inert, soluble in NMR solvents, and easily removable. Also, the highly shielded methyl protons in TMS yield an intense...
Proton (¹H) NMR: Chemical Shift
Absorption signals of all the protium nuclei...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
¹H NMR of Labile Protons: Deuterium (²H) Substitution
¹H NMR of Labile Protons: Temporal Resolution
The –OH proton in alcohols typically appears in the range of δ 2 to 5 ppm but can vary depending on the specific...