Analysis of objective quality metrics in computed tomography images affected by metal artifacts
Yakdiel Rodriguez-Gallo1, Ruben Orozco-Morales1, Marlen Perez-Diaz1
1Departamento de Control Automático, Universidad Central 'Marta Abreu' de Las Villas, Santa Clara, Cuba.
Abstract:
Image quality (IQ) assessment plays an important role in the medical world. New methods to evaluate image quality have been developed, but their application in the context of computer tomography is yet limited. In this paper the performance of fifteen well-known full reference (FR) IQ metrics is compared with human judgment using images affected by metal artifacts and processed with metal artifact reduction methods from a phantom. Five region of interest with different sizes were selected. IQ was evaluated by seven experienced radiologists completely blinded to the information. To measure the correlation between FR-IQ, and the score assigned by radiologists non-parametric Spearman rank-order correlation coefficient and Kendall's Rank-order Correlation coefficient were used; so as root mean square error and the mean absolute error to measure the prediction accuracy. Cohen's kappa was employed with the purpose of assessing inter-observer agreement. The metrics GMSD, IWMSE, IWPSNR, WSNR and OSS-PSNR were the best ranked. Inter-observer agreement was between 0.596 and 0.954, with p<0.001 in all study. The objective scores predicted by these methods correlate consistently with the subjective evaluations. The application of this metrics will make possible a better evaluation of metal artifact reduction algorithms in future works.
More Related Videos
08:39Longitudinal Micro-Computed Tomography Image Analysis for User-Defined Region of Interest in Critical-Sized Bone Defects
Published on: June 24, 2025
10:10Three-Dimensional Particle Shape Analysis Using X-ray Computed Tomography: Experimental Procedure and Analysis Algorithms for Metal Powders
Published on: December 4, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...