Cortical Bone Assessment Using Ultrasonic Guided Waves: A Reproducibility Study in a Healthy Population
Published on: January 31, 2025
Interobserver Agreement and Correlation of an Automated Algorithm for B-Line Identification and Quantification With
Christopher L Moore1, Jing Wang2, Anthony J Battisti2
1Department of Emergency Medicine, Yale School of Medicine, New Haven, CT, USA.
Objectives:
B-lines are ultrasound artifacts that can be used to detect a variety of pathologic lung conditions. Computer-aided methods to detect and quantify B-lines may standardize quantification and improve diagnosis by novice users. We sought to test the performance of an automated algorithm for the detection and quantification of B-lines in a handheld ultrasound device (HHUD).
Methods:
Ultrasound images were prospectively collected on adult emergency department patients with dyspnea. Images from the first 124 patients were used for algorithm development. Clips from 80 unique subjects for testing were randomly selected in a predefined proportion of B-lines (0 B-lines, 1-2 B-lines, 3 or more B-lines) and blindly reviewed by five experts using both a manual and reviewer-adjusted process. Intraclass correlation coefficient (ICC) and weighted kappa were used to measure agreement, while an a priori threshold of an ICC (3,k) of 0.75 and precision of 0.3 were used to define adequate performance.
Results:
ICC between the algorithm and manual count was 0.84 (95% confidence interval [CI] 0.75-0.90), with a precision of 0.15. ICC between the reviewer-adjusted count and the algorithm count was 0.94 (95% CI 0.90-0.96), and the ICC between the manual and reviewer-adjusted counts was 0.94 (95% CI 0.90-0.96). Weighted kappa was 0.72 (95% CI 0.49-0.95), 0.88 (95% CI 0.74-1), and 0.85 (95% CI 0.89-0.96), respectively.
Conclusions:
This study demonstrates a high correlation between point-of-care ultrasound experts and an automated algorithm to identify and quantify B-lines using an HHUD. Future research may incorporate this HHUD in clinical studies in multiple settings and users of varying experience levels.
More Related Videos
16:01An Experimental Protocol for Assessing the Performance of New Ultrasound Probes Based on CMUT Technology in Application to Brain Imaging
Published on: September 24, 2017
08:52Software-Assisted Quantitative Measurement of Osteoarthritic Subchondral Bone Thickness
Published on: March 18, 2022
Related Concept Videos
Imaging Studies II: Ultrasonography
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):