1Biomedical Imaging Research Group (GIBI230) at La Fe Health Research Institute and Imaging La Fe node at Distributed Network for Biomedical Imaging (ReDIB) Unique Scientific and Technical Infrastructures (ICTS), La Fe University and Polytechnic Hospital, Av. Fernando Abril Martorell, 106, Torre E, Planta 0, 46026, Valencia, Spain. marti_lui@gva.es.
Background:
Indirect 1H-magnetic resonance (MR) imaging of 17O-labelled water allows imaging in vivo dynamic changes in water compartmentalisation. Our aim was to describe the feasibility of indirect 1H-MR methods to evaluate the effect of H217O on the MR relaxation rates by using conventional a 3-T equipment and voxel-wise relaxation rates.
Methods:
MR images were used to calculate the R1, R2, and R2* relaxation rates in phantoms (19 vials with different H217O concentrations, ranging from 0.039 to 5.5%). Afterwards, an experimental animal pilot study (8 rats) was designed to evaluate the in vivo relative R2 brain dynamic changes related to the intravenous administration of 17O-labelled water in rats.
Results:
There were no significant changes on the R1 and R2* values from phantoms. The R2 obtained with the turbo spin-echo T2-weighted sequence with 20-ms echo time interval had the higher statistical difference (0.67 s-1, interquartile range 0.34, p < 0.001) and Spearman correlation (rho 0.79). The R2 increase was adjusted to a linear fit between 0.25 and 5.5%, represented with equation R2 = 0.405 concentration + 0.3215. The highest significant differences were obtained for the higher concentrations (3.1-5.5%). The rat brain MR experiment showed a mean 10% change in the R2 value after the H217O injection with progressive normalisation.
Conclusions:
Indirect 1H-MR imaging method is able to measure H217O concentration by using R2 values and conventional 3-T MR equipment. Normalised R2 relative dynamic changes after the intravenous injection of a H217O saline solution provide a unique opportunity to map water pathophysiology in vivo, opening the analysis of aquaporins status and modifications by disease at clinically available 3-T proton MR scanners.
Related Concept Videos
2D NMR: Heteronuclear Single-Quantum Correlation Spectroscopy (HSQC)
2D NMR: Overview of Heteronuclear Correlation Techniques
2D NMR: Overview of Homonuclear Correlation Techniques
COSY90 is the standard two-dimensional (2D) COSY experiment that...
Imaging Studies IV: Magnetic Resonance Imaging
¹H NMR of Labile Protons: Deuterium (²H) Substitution
Imaging Studies III: Computed Tomography