Adapting a continuous flow cryostat and a plate DAC to do high pressure Raman experiments at low temperatures
Smita Gohil1, Shankar Ghosh1, Satej Tare2
1Department of Condensed Matter Physics and Materials Science, Tata Institute of Fundamental Research, Colaba, Mumbai 400005, India.
Abstract:
We present a method for modifying a continuous flow cryostat and a steel plate DAC (Diamond Anvil Cell) to perform high pressure micro-Raman experiments at low temperatures. Despite using a steel DAC with a lower specific heat capacity (∼335 J/kg K), this setup can routinely perform high pressure (∼10 GPa) measurements at temperatures as low as 26 K. This adaptation is appropriate for varying the temperature of the sample while keeping it at a constant pressure. We determined that the temperature variation across the sample chamber is about 1 K using both direct temperature measurements and finite element analysis of the heat transport across the DAC. We present Raman spectroscopy results on elemental selenium at high pressures and low temperatures using our modified setup.
Related Concept Videos
Constant Pressure Calorimetry
Raman Spectroscopy Instrumentation: Overview
The monochromatic laser source, typically using visible or near-infrared radiation, generates a highly focused beam of light. This light interacts with the molecules of the sample, scattering some of the light. Liquid and gaseous samples are usually tested in ordinary glass capillaries, while solids can be analyzed as powders packed in capillaries or as potassium...