Purpose:
To use MC simulations and phantom measurements to investigate the dosimetry of a kilovoltage x-ray beam from an IR fluoroscope to deliver low-dose (0.3-1.0 Gy) radiotherapy to the lungs.
Materials And Methods:
PENELOPE was used to model a 125 kV, 5.94 mm Al HVL x-ray beam produced by a fluoroscope. The model was validated through depth-dose, in-plane/cross-plane profiles and absorbed dose at 2.5-, 5.1-, 10.2- and 15.2-cm depths against the measured beam in an acrylic phantom. CT images of an anthropomorphic phantom thorax/lungs were used to simulate 0.5 Gy dose distributions for PA, AP/PA, 3-field and 4-field treatments. DVHs were generated to assess the dose to the lungs and nearby organs. Gafchromic film was used to measure doses in the phantom exposed to PA and 4-field treatments, and compared to the MC simulations.
Results:
Depth-dose and profile results were within 3.2% and 7.8% of the MC data uncertainty, respectively, while dose gamma analysis ranged from 0.7 to 1.0. Mean dose to the lungs were 1.1-, 0.8-, 0.9-, and 0.8- Gy for the PA, AP/PA, 3-field, and 4-field after isodose normalization to cover ∼ 95% of each lung volume. Skin dose toxicity was highest for the PA and lowest for the 4-field, and both arrangements successfully delivered the treatment on the phantom. However, the dose distribution for the PA was highly non-uniform and produced skin doses up to 4 Gy. The dose distribution for the 4-field produced a uniform 0.6 Gy dose throughout the lungs, with a maximum dose of 0.73 Gy. The average percent difference between experimental and Monte Carlo values were -0.1% (range -3% to +4%) for the PA treatment and 0.3% (range -10.3% to +15.2%) for the 4-field treatment.
Conclusion:
A 125 kV x-ray beam from an IR fluoroscope delivered through two or more fields can deliver an effective low-dose radiotherapy treatment to the lungs. The 4-field arrangement not only provides an effective treatment, but also significant dose sparing to healthy organs, including skin, compared to the PA treatment. Use of fluoroscopy appears to be a viable alternative to megavoltage radiation therapy equipment for delivering low-dose radiotherapy to the lungs.
More Related Videos
07:57Positron Emission Tomography-based Dose Painting Radiation Therapy in a Glioblastoma Rat Model using the Small Animal Radiation Research Platform
Published on: March 24, 2022
08:54Treatment of Liver Metastases Using an Internal Target Volume Method for Stereotactic Body Radiotherapy
Published on: May 8, 2018
Related Concept Videos
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...