Intra-patient comparison of 3D and 2D magnetic resonance elastography techniques for assessment of liver stiffness

Roberta Catania1, Camila Lopes Vendrami2, Bradley D Bolster3

  • 1Department of Radiology, Northwestern University Feinberg School of Medicine, 676 N. Saint Clair Street, Suite 800, Chicago, IL, 60611, USA. roberta.catania@northwestern.edu.

Abstract

Insights

Three-dimensional magnetic resonance elastography (3D MRE) offers more reproducible hepatic stiffness measurements than 2D techniques. This advanced method provides larger parenchymal areas for comprehensive liver evaluation with reduced artifact susceptibility.

Area of Science:

  • Medical Imaging
  • Hepatology
  • Biophysics

Background:

  • Liver stiffness measurement (LSM) is crucial for diagnosing and monitoring liver diseases.
  • Conventional 2D magnetic resonance elastography (MRE) techniques have limitations in reproducibility and artifact susceptibility.
  • Advancements in MRE aim to improve accuracy and comprehensiveness in assessing hepatic stiffness.

Purpose of the Study:

  • To evaluate the performance of 3D MRE using spin-echo echo-planar imaging (seEPI) for hepatic stiffness assessment.
  • To compare 3D MRE (seEPI) with conventional 2D gradient-recalled echo (GRE) and 2D seEPI sequences.
  • To assess the impact of region of interest (ROI) placement and abbreviated analysis on 3D MRE measurements.

Main Methods:

  • Retrospective analysis of 57 liver MRE examinations using 2D GRE, 2D seEPI, and 3D seEPI sequences.
  • Elastograms analyzed with both "curated" (artifact-avoiding) and "non-curated" ROIs.
  • Liver stiffness measurements (LSMs) calculated, including abbreviated LSMs from 4 slices for 3D MRE.
  • Statistical analysis using univariate tests with a significance level of p < 0.05.

Main Results:

  • 3D MRE provided significantly larger ROI surface areas compared to 2D techniques.
  • 3D LSMs using curated and non-curated ROIs showed no clinically significant difference (<0.1 kPa mean difference).
  • Abbreviated 3D LSMs demonstrated excellent correlation (r=0.9, p<0.001) and no significant difference compared to full 3D MRE LSMs (p=0.927).

Conclusions:

  • 3D MRE enables more reproducible hepatic stiffness measurements due to reduced susceptibility to artifacts.
  • The technique allows for evaluation of larger liver parenchymal areas, leading to a more comprehensive assessment.
  • Abbreviated 3D MRE analysis provides reliable and correlated results with full 3D MRE, enhancing clinical utility.