SIR-Spheres Activity Measurements Reveal Systematic Miscalibration
Stephen A Graves1,2,3, Molly Martin4, Ashok Tiwari4,5
1Department of Radiology, University of Iowa, Iowa City, Iowa; stephen-a-graves@uiowa.edu.
Abstract:
The purpose of this work was to perform an independent and National Institute of Standards and Technology-traceable activity measurement of 90Y SIR-Spheres (Sirtex). γ-spectroscopic measurements of the 90Y internal pair production decay mode were made using a high-purity germanium detector. Methods: Measured annihilation radiation detection rates were corrected for radioactive decay during acquisition, dead time, source attenuation, and source geometry effects. Detection efficiency was determined by 2 independent and National Institute of Standards and Technology-traceable methods. Results: Measured SIR-Spheres vials (n = 5) contained more activity than specified by the manufacturer calibration; on average, the ratio of measured activity to calibrated was 1.233 ± 0.030. Activity measurements made using 2 distinct efficiency calibration methods agreed within 1%. Conclusion: The primary SIR-Spheres activity calibration appears to be a significant underestimate of true activity.
More Related Videos
Related Concept Videos
Random and Systematic Errors
Uncertainty in Measurement: Accuracy and Precision
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Uncertainty in Measurement: Reading Instruments
Common Leveling Mistakes and Errors
Measurement: Standard Units