Clinical Application of a High Sensitivity BGO PET/CT Scanner: Effects of Acquisition Protocols and Reconstruction
Elena De Ponti1, Cinzia Crivellaro2, Sabrina Morzenti1
1Department of Medical Physics, San Gerardo Hospital, ASST - Monza, Via Pergolesi 33, Monza (MB) 20900, Italy.
Aims:
The aim of this retrospective study was to investigate SUVs variability with respect to lesion size, administered dose, and reconstruction algorithm.
Background:
SUVmax and SUVpeak are influenced by technical factors as count statistics and reconstruction algorithms.
Objective:
To fulfill the aim, we evaluated the SUVs variability with respect to lesion size, administered dose, and reconstruction algorithm (ordered - subset expectation maximization plus point spread function option - OSEM+PSF, regularized Bayesian Penalized Likelihood - BPL) in a 5 - rings BGO PET/CT scanner.
Methods:
Discovery IQ scanner (GE Healthcare, Milwaukee, Wisconsin, US) was used for list mode acquisition of 25 FDG patients, 12 injected with 3.7 MBq/kg (Standard Dose protocol - SD) and 13 injected with 1.8 MBq/kg (Low Dose protocol - LD). Each acquisition was reconstructed at different time/FOV with both OSEM+PSF algorithm and BPL using seven different beta factors. SUVs were calculated in 70 lesions and analysed in function of time/FOV and Beta. Image quality was evaluated as a coefficient of variation of the liver (CV - liver).
Results:
SUVs were not considerably affected by time/FOV. However, SUVs were influenced by beta: differences were higher in small lesions (37% for SUVmax, 15% for SUVpeak) compared to larger ones (14% and 6%). CV - liver ranged from 6% with Beta-500 (LD and SD) to 13% with Beta- 200 (LD). CV - liver of BPL with Beta-350 (optimized for clinical practice in our institution) in LD was lower than CV - liver of OSEM+PSF in SD.
Conclusion:
When a high sensitivity 5 - rings BGO PET/CT scanner is used with the same reconstruction algorithm, quantification by means of SUVmax and SUVpeak is a robust standard compared to the activity and scan duration. However, both SUVs and image quality are influenced by reconstruction algorithms and the related parameters should be considered to obtain the best compromise between detectability, quantification, and noise.
More Related Videos
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET