Magnet Fracture within a Magnetically Controlled Growing Rod: A Case Report of a New Mechanism of Failure
Conor S Jones1, Paul Rushton2, Michael Hutton1
1Department of Spine, Exeter Spine Unit, Royal Devon and Exeter NHS Foundation Trust, Exeter.
Introduction:
Magnetically controlled growing rods (MCGRs) have been widely adopted in the management of early-onset scoliosis since they were first described in 2012. Recent reports have highlighted concerns around their safety. To date, little is understood about the risk factors and modes of failure in these devices.
Case Report:
We report a novel mechanism of device failure in a 14-year-old patient following multiple revisions of MCGRs.Clinically, there was no evidence of device failure and the MCGRs appeared radiologically intact. Explantation analysis revealed multiple compromised/non-functional components. A previously undocumented phenomenon of complete magnet fracture was also seen.
Conclusion:
The absence of clinical or radiological features of device failure in this case makes the findings of great concern. Given the relative paucity of high-quality evidence surrounding the use of MCGRs, we support calls for urgent comparative studies and further investigation of risk factors for device failure.