Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Accuracy of Photorefraction and Wavefront-Based Autorefraction in Children Under 3 Years of Age
Caner Kara1, Ikbal Seza Petricli1
1Department of Ophthalmology, Etlik Zübeyde Hanım Women's Health Training and Research Hospital, Ankara, Turkey.
Insights
Photorefraction and wavefront-based autorefraction underestimate refractive error in young children. Wavefront autorefraction with cycloplegia may be an alternative to cycloplegic retinoscopy for pediatric eye exams.
Area of Science:
- Ophthalmology
- Pediatric Optometry
- Vision Science
Background:
- Accurate refractive error assessment in children under 3 is crucial for early detection and management of visual impairments.
- Traditional methods like cycloplegic retinoscopy are considered the gold standard but can be challenging in very young children.
- Non-cycloplegic autorefraction and photorefraction offer potential alternatives for faster screening.
Purpose of the Study:
- To evaluate the diagnostic accuracy of photorefraction (Spot Vision Screener) and wavefront-based autorefraction (SureSight Vision Screener) compared to cycloplegic retinoscopy in children under 3 years old.
- To assess the performance of these devices before and after cycloplegia.
Main Methods:
- A cross-sectional study included 305 children under 3 years with refractive errors.
- Measurements of spherical equivalent (SE) and Jackson cross-cylinder values were taken using Spot Vision Screener and SureSight Vision Screener, both before and after cycloplegia.
- Results were compared against cycloplegic retinoscopy.
Main Results:
- Both devices significantly underestimated SE values before cycloplegia compared to retinoscopy (p<0.001).
- After cycloplegia, SureSight Vision Screener showed no significant difference compared to retinoscopy (p=0.66).
- Spot Vision Screener, even after cycloplegia, yielded significantly higher SE values than retinoscopy (p<0.001).
Conclusions:
- Both photorefraction and autorefraction devices tend to underestimate refractive error in this age group.
- Wavefront-based autorefraction (SureSight) with cycloplegia shows promise as a viable alternative to cycloplegic retinoscopy for pediatric refractive assessment.
- Further studies are needed to confirm the utility of wavefront autorefraction in this population.
Objectives:
This study was designed to assess the accuracy of photorefraction and wavefront-based autorefraction in children under 3 years of age.
Methods:
A cross-sectional study was performed with children who had no ocular abnormalities other than a refractive error and were less than 3 years old. Spot Vision Screener (Welch Allyn Inc., Skaneateles Falls, NY, USA) was used for photorefraction and SureSight Vision Screener (Welch Allyn, Skaneateles Falls, NY) was used for autorefraction. The measurements were performed with both devices before and after cycloplegia. The results were compared with those obtained using cycloplegic retinoscopy. The assessments were performed in terms of spherical equivalent (SE) values and Jackson cross-cylinder power values (J0 and J45).
Results:
A total of 305 children with a mean age of 15.1±10.4 months were included. The mean SE values measured with the Spot Vision and SureSight before cycloplegia were 0.43±0.75 D and 0.90±0.86 D, respectively. Following cycloplegia, the mean SE values were 1.80±1.25 D and 1.55±1.40 D, respectively. The mean SE value obtained with cycloplegic retinoscopy was 1.53±1.34 D. The mean SE values measured with both devices prior to cycloplegia were significantly lower than that measured with cycloplegic retinoscopy (p<0.001). No significant difference was observed between the cycloplegic SE values obtained with SureSight and retinoscopy (p=0.66). Spot Vision yielded significantly higher cycloplegic SE values compared with cycloplegic retinoscopy (p<0.001).
Conclusion:
Both devices underestimated SE values in comparison with cycloplegic retinoscopy in children under 3 years of age. Wavefront-based autorefraction performed with cycloplegia may be an alternative to cycloplegic retinoscopy, but needs confirmation.