Objectives:
This study characterized a violet LED light (V-LED; bright max whitening) tooth whitening device and evaluated its efficacy on stained enamel compared to hydrogen peroxide (HP).
Materials And Methods:
Characterization of the V-LED beam profile was performed using a laser beam-profiler. The irradiance was measured throughout an exposure cycle at 0- and 8-mm distances using an integrating sphere and a spectral radiometer. Bovine enamel/dentin blocks stained with black tea (BT), cigarette smoke (CS), or without staining (CONT) were subjected to V-LED or 40% HP (n = 10/group). Color parameters (ΔL, Δa, Δb, and ΔE00) were measured using a digital spectrophotometer. Light transmission was estimated through 1-mm-thick bovine enamel slices (n = 5). ΔL, Δb, ΔE00, and irradiance were analyzed by two-way ANOVAs and Tukey's tests, Δa by Kruskal-Wallis and Mann-Whitney tests, and light transmission by t-test (α = 5%).
Results:
Heterogeneous beam distribution was observed for the emitting V-LED chips. After 20 sequential exposures, irradiance levels were reduced 25-50%, regardless of the distance from V-LED. Localized irradiance values were statistically different between beam locations and different distances from the target. V-LED produced lower ΔE00, ΔL, Δa, and Δb values than HP for CONT and BT, with no differences for CS. Light transmittance decreased approximately 98% through 1-mm thick enamel.
Conclusions:
V-LED irradiance was heterogeneous and decreased throughout the exposure cycles and was also greatly reduced with increasing tip distance. V-LED produced a significantly lower whitening effect on BT and control teeth.
Clinical Relevance:
This study contributes to the knowledge of V-LED and its clinical use.