Comparison of 360 degrees and 180 degrees data collection in SPECT imaging
1Clinic of Nuclear Medicine and Oncology, Clinical Hospital Dr M Stojanovic, Zagreb, SR Croatia, Yugoslavia.
Physics in Medicine and Biology
|November 1, 1987
Abstract:
The problem of using 360 degrees or 180 degrees data sampling techniques in transaxial SPECT imaging is still to be solved. A theoretical point source study for an ideal response detector has shown, for objects which are close enough to the origin of the reconstructed area, that there are significant differences between sections obtained by different sampling methods. A computer simulation study of line sources in a homogeneous attenuated medium has confirmed the results of clinical studies in which significant image distortion has been observed in 180 degrees sections but not in 360 degrees reconstructed images.