Susceptibility artifacts induced by crowns of different materials with prepared teeth and titanium implants in
Xiaomeng Gao1,2, Qianbing Wan2, Qingping Gao3
1Department of Stomatology, Xiangya Hospital, Central South University, 87 Xiangya Road, Changsha, 410008, Hunan, China.
Abstract:
This study aimed to investigate the artifacts induced by crowns composed of different materials with prepared teeth and titanium implants. Resin, metal-ceramic, ceramic and zirconia crowns were fabricated and placed onto the prepared teeth on a human cadaver head or titanium implants with prosthesis abutments on a dry human mandible. The samples were scanned on a 1.5 T MRI apparatus, and artifact areas were defined as the signal intensity and signal loss adjacent to the prosthesis and measured by a threshold tool with ImageJ2x. Data were analyzed using SPSS 22.0. Resin, ceramic, zirconia, and precious metal-ceramic crowns barely produced artifacts on the cadaver skull (p > 0.999). By contrast, pure Ti and nonprecious metal-ceramic crowns created significant artifacts (p < 0.001). The average artifacts reduction of double Au-Pt and Ag-Pd metal-ceramic crowns combined with titanium implants and abutments was 79.49 mm2 (p < 0.001) and 74.17 mm2 (p < 0.001) respectively, while artifact areas were increased in double Co-Cr and Ni-Cr metal-ceramic crowns by 150.10 mm2 (p < 0.001) and 175.50 mm2 (p < 0.001) respectively. Zirconia, ceramic and precious metal-ceramic crowns induce less MRI artifacts after tooth preparation while precious metal-ceramic crowns alleviate artifacts in combination with titanium implants.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System IV: CMRI
Imaging Studies IV: Magnetic Resonance Imaging