Agreement of ocular response analyzer cornea compensated IOP with corvis ST biomechanical IOP following Femtosecond
Ramin Salouti1,2, Reza Razeghinejad3, Gholamreza Eslami1
1Department of Ophthalmology, School of Medicine, Shiraz University of Medical Sciences, Shiraz, Iran.
Objectives:
To compare intraocular pressure (IOP) measurement by ORA-IOPcc and Corvis-bIOP after femtosecond laser-assisted LASIK (FS-LASIK).
Methods:
In this prospective cohort study, 56 eyes from 56 consecutive patients scheduled for FS-LASIK were enrolled. All patients had IOP measurement with ORA and Corvis ST by two blinded independent expert examiners. IOP examinations were conducted between 8 and 11 A.M. Data were collected at baseline and 3 months after FS-LASIK.
Results:
The mean age of the participants was 29.1 ± 6.3 years, and 42 (75%) were female. The average of central corneal thickness (CCT) decreased from 537 ± 23 µm at baseline to 458 ± 31 µm after FS-LASIK. The mean postoperative change of IOP was 0.0 ± 2.1 for bIOP and -2.5 ± 3.2 mmHg for IOPcc. The corresponding 95% limits of agreement (LoA) was -4.1 to 4.1 mmHg and -3.8 to 8.8 mmHg, respectively. Both methods showed no significant correlation between ∆IOP and ∆CCT. The 95% LoA between bIOP and IOPcc after FS-LASIK was -4.8 to 9.1 mmHg.
Conclusions:
Compared to the ORA-IOPcc, the Corvis-bIOP showed less variation after FS-LASIK and might be a more appropriate choice for measuring IOP in this condition. The agreement of bIOP vs. IOPcc after FS-LASIK is below the clinically acceptable level, and the two methods could not be regarded as interchangeable.