Related Experiment Video
Updated: Oct 6, 2025

Reverse Total Shoulder Arthroplasty
Published on: July 5, 2011
Mid-term outcomes of a stemless ceramic head anatomic total shoulder replacement
Maciej J K Simon1,2, Jennifer A Coghlan3,4, Jeff Hughes5
1Melbourne Shoulder and Elbow Centre, 1/80 Beach Road, Sandringham, VIC, 3191, Australia. maciej.simon@uksh.de.
This study shows that stemless anatomic shoulder replacement with a ceramic head and polyethylene glenoid component offers excellent mid-term clinical and radiographic outcomes for osteoarthritis. Patients experienced significant improvements in range of motion, pain reduction, and overall satisfaction.
Area of Science:
- Orthopedic Surgery
- Biomaterials Science
- Clinical Research
Background:
- Stemless anatomic shoulder replacement (aTSR) shows promise for treating glenohumeral osteoarthritis.
- Ceramic-on-polyethylene articulations may reduce wear particle generation compared to metal-on-polyethylene.
- Mid-term clinical and radiographic outcomes of this specific implant combination require evaluation.
Purpose of the Study:
- To assess the clinical and radiographic results of stemless aTSR using a ceramic head and polyethylene glenoid component.
- To evaluate patient-reported outcomes and functional improvements at mid-term follow-up.
- To determine the incidence of complications such as glenoid radiolucency and osteolysis.
Main Methods:
- Prospective study of 92 patients undergoing stemless aTSR for glenohumeral osteoarthritis.
- Utilized a stemless humeral component with a ceramic head and a cemented polyethylene glenoid.
- Clinical outcomes (ASES, Constant, SPADI, DASH, VAS pain, satisfaction, ROM) assessed at 2 and 5 years.
- Radiographic assessment included glenoid radiolucent lines and humeral osteolysis.
Main Results:
- At 5-year follow-up (n=74), active elevation improved from 91.3° to 151.1° (p<0.001).
- Significant improvements noted in ASES (41.6 to 94.3), SPADI (62.9 to 4.3), and VAS pain (5.6 to 0.4).
- 96% patient satisfaction; 62 cases showed no glenoid radiolucent lines; one revision for glenoid loosening.
Conclusions:
- Stemless anatomic shoulder replacement with a ceramic head and polyethylene glenoid demonstrates favorable 5-year clinical and radiographic outcomes.
- The implant combination provides significant functional improvement and high patient satisfaction.
- Low rates of radiographic complications suggest good implant stability and durability.