Lateral Resolution of a Commercial Optical Coherence Tomography Instrument
Richard F Spaide1, Tilman Otto2, Sophie Caujolle2
1Vitreous Retina, Macula Consultants of New York, New York, New York, USA.
Purpose:
The lateral resolution of an optical coherence tomography (OCT) instrument was considered to be equal to the illumination spot size on the retina. To evaluate the potential lateral resolution of the Spectralis OCT, an instrument calculated to have a 14 µm resolution.
Methods:
The lateral point spread function (PSF) was evaluated using diamond abrasive powder 0 to 1 µm in diameter in silicone elastomer and a validated target with 800 nm FeO particles in urethane. The amplitude transfer function was calculated from human OCT images. Finally, resolution was measured using the 1951 USAF target.
Results:
Measurement of the lateral PSF from 1215 diamond particle images yielded a full-width half maximum (FWHM) to be 5.11 µm and for 732 FeO particles, 4.9 µm. From the amplitude transfer function, the FWHM of the diffraction limited PSF was calculated to be 5.0 µm. The USAF target imaging showed a lateral resolution of 4.6 µm.
Conclusions:
Although a calculation of the spot size of the illumination beam was reported in the past as the lateral resolution of the OCT instrument, the actual lateral resolution is better by a factor of at least 2.5 times. The clinically used A-scan spacing was derived from the calculated, and not the true resolution, and results in under sampling. This set of findings likely apply to all commercial clinical instruments.
Translational Relevance:
The scan density parameters of past and present commercial OCT instruments were based on earlier translational concepts, which now appear to have been incorrect.
More Related Videos
12:54Simultaneous Brightfield, Fluorescence, and Optical Coherence Tomographic Imaging of Contracting Cardiac Trabeculae Ex Vivo
Published on: October 2, 2021
09:56Automated 3D Optical Coherence Tomography to Elucidate Biofilm Morphogenesis Over Large Spatial Scales
Published on: August 21, 2019
Related Concept Videos
Imaging Biological Samples with Optical Microscopy
In optical microscopy, the specimen to be viewed is placed on a glass slide and clipped on the stage...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Confocal Fluorescence Microscopy
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...