Validation of Continuous Noninvasive Blood Pressure Monitoring Using Error Grid Analysis
Takashi Juri1, Koichi Suehiro1, Rieko Kanematsu1
1From the Department of Anesthesiology, Osaka City University Graduate School of Medicine, Osaka, Japan.
Background:
Error grid analysis was recently proposed to compare blood pressure obtained by 2 measurement methods. This study aimed to compare continuous noninvasive blood pressure (CNBP) with invasive blood pressure (IBP) using the error grid analysis and investigate the confounding risk factors attributable to the differences between CNBP and IBP.
Methods:
Sixty adult patients undergoing general anesthesia were prospectively enrolled. Simultaneous comparative data regarding CNBP and IBP were collected. The Bland-Altman analysis was conducted to compare CNBP and IBP for systolic blood pressure (SBP) and mean blood pressure (MBP; acceptable accuracy: mean bias <5 mm Hg; standard deviation <8 mm Hg). The clinical relevance of the discrepancies between CNBP and IBP was evaluated by the error grid analysis, which classifies the differences into 5 zones from "no risk" (A) to "dangerous risk" (E). Additionally, an ordinal logistic regression analysis was performed to evaluate the relationship between the risk zones for MBP, classified by the error grid analysis and covariates of interest.
Results:
A total of 10,663 pairs of CNBP/IBP were finally analyzed. The Bland-Altman analysis showed an acceptable accuracy with a bias of -3.3 ± 5.6 mm Hg for MBP but a poor accuracy with a bias of 5.4 ± 10.5 mm Hg for SBP. The error grid analysis showed the proportions of zones A to E as 96.7%, 3.2%, 0.1%, 0%, and 0% for SBP, respectively, and 72.0%, 27.9%, 0.1%, 0%, and 0% for MBP, respectively. The finger cuff missed 23.9% of epochs when SBP <90 mm Hg and 55.3% of epochs when MBP <65 mm Hg. The ordinal logistic regression analysis revealed that older age (adjusted odds ratio for decade: 1.54, 95% confidence interval [CI], 1.15-2.08; P = .004) and length of time from the initiation of finger cuff inflation (adjusted odds ratio for 60 minutes: 1.40, 95% CI, 1.13-1.73; P = .002) were significant factors of being in a more dangerous zone of the error grid.
Conclusions:
The error grid analysis revealed the larger clinical discrepancy between CNBP and IBP in MBP compared with that in SBP. Old age and longer finger cuff inflation time were significant factors of being in a more dangerous zone of the error grid, which could affect the hemodynamic management during surgery.
Related Concept Videos
Errors occurring during blood pressure monitoring
Several factors...
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pre-Procedural Guidelines for Assessing Blood Pressure
Measurement of Blood Pressure
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement