Percutaneous decannulation of extracorporeal membrane oxygenation using a plug-based closure device
Magnus Dalén1,2, Magnus Settergren2,3, Mikael Kastengren2,3
1Department of Cardiothoracic Surgery, Karolinska University Hospital, Stockholm, Sweden.
Background:
There is limited experience of using the MANTA plug-based vascular closure device for percutaneous arterial closure of the femoral artery after venoarterial extracorporeal membrane oxygenation.
Objectives:
To study femoral artery complications and need for subsequent vascular interventions after percutaneous decannulation of venoarterial extracorporeal membrane oxygenation (VA ECMO) using the MANTA plug-based vascular closure device.
Methods:
We studied 34 consecutive patients who underwent percutaneous decannulation of VA ECMO using the MANTA device. Primary outcomes were conversion to surgical cutdown of the groin at decannulation (immediate) or later. Secondary outcomes were type of vascular complication necessitating conversion to surgical cutdown of the groin.
Results:
Six (17.7%) patients had to undergo immediate (n = 3) or late (n = 3) conversion to surgical cutdown of the groin. Of these, three were owing to occlusion of the common femoral artery resulting in insufficient distal perfusion and three owing to bleeding or pseudoaneurysm. The mechanism of failure was complete intravascular deployment of the MANTA device in three patients, incomplete MANTA sealing of the arteriotomy in one patient, MANTA-unrelated thrombotic occlusion in one patient, and unknown in one patient. Surgical cut-down was typically performed with concomitant catheter thrombectomy with or without patch reconstruction of the artery.
Conclusion:
Percutaneous decannulation of VA ECMO using the MANTA VCD was feasible but a substantial number of patients needed to be converted to unplanned surgical repair, owing to either closure site-located stenosis/occlusion or bleeding. If suboptimal MANTA positioning is suspected, a low threshold for conversion to surgical cutdown of the groin is recommended.
Related Concept Videos
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Extracorporeal Removal of Drugs: Peritoneal Dialysis and Hemodialysis
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration