Related Experiment Video
Updated: Oct 5, 2025

Effects of Mechanical Methods Used in Peri-implantitis Treatment on Implant Surface Decontamination and Roughness
Published on: March 14, 2025
Stability of dental implants with sandblasted and acid-etched (SLA) and modified (SLActive) surfaces during the
Gulsum Sayin Ozel1, Ozgur Inan2, Asli Secilmis Acar3
1Department of Prosthodontics, Istanbul Medipol University, Istanbul, Turkey.
Abstract:
Background. The surface properties of implants are effective factors for increasing the osseointegration and activity of osteoprogenitor cells. This study compared the stability of dental implants with sandblasted and acid-etched (SLA) and modified surfaces (SLActive) using the resonance frequency analysis (RFA). Methods. In a split-mouth design, 50 dental implants with either SLA surface properties (n=25) or modified (SLActive) surface properties (n=25) were placed in the mandibles of 12 patients with a bilateral posterior edentulous area. Implant stability was measured using RFA (Osstell) at implant placement time and every week for 1, 2, and 3 months before the conventional loading time. Results. One week following the implantation, implant stability increased from 70 to 77.67 for SLA and from 71.67 to 79 for SLActive (P < 0.05). Stability improved each week except in the 4th week in SLActive surface measurements. No significant differences were observed between the groups at 2 and 3 months (P > 0.05). Conclusions. For both implant surfaces, increased stability was observed over time, with no significant differences between the groups.

