A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
The "Defibrillation Testing, Why Not?" survey. Testing of subcutaneous and transvenous defibrillators in the Italian
Federico Migliore1, Stefano Viani2, Matteo Ziacchi3
1Department of Cardiac, Thoracic Vascular Sciences and Public Health University of Padova, Italy.
Background:
Defibrillation testing (DT) can be omitted in patients undergoing transvenous implantable cardioverter-defibrillator (T-ICD) implantation, but it is still recommended for patients at risk for a high defibrillation threshold and for ICD generator changes. Moreover, DT is still recommended on implantation of subcutaneous ICD (S-ICD). The aim of the present survey was to analyze the current practice of DT during T-ICD and S-ICD implantations.
Methods:
In March 2021, an ad hoc questionnaire on the current performance of DT and the standard practice adopted during testing was completed at 72 Italian centers implanting S-ICD and T-ICD.
Results:
48 (67%) operators reported never performing DT during de-novo T-ICD implantations, while no operators perform it systematically. The remaining respondents perform it for patients at risk for a high defibrillation threshold. DT is never performed at T-ICD generator change. At the time of de-novo S-ICD implantation, DT is never performed by 9 (13%) operators and performed systematically by 48 (66%). The remaining operators frequently omit DT in patients with more severe systolic dysfunction. DT is not performed at S-ICD generator change by 92% of operators. DT is conducted by delivering a first shock energy of 65 J by 60% of operators, while the remaining 40% test lower energy values.
Conclusions:
In current clinical practice, most operators omit DT at T-ICD implantation, even when still recommended in the guidelines. DT is also frequently omitted at S-ICD implantation, and a wide variability exists among operators in the procedures followed during DT.
More Related Videos
08:00Reduction of Iatrogenic Atrial Septal Defects with an Anterior and Inferior Transseptal Puncture Site when Operating the Cryoballoon Ablation Catheter
Published on: June 15, 2015
05:36Standardized Model of Ventricular Fibrillation and Advanced Cardiac Life Support in Swine
Published on: January 30, 2020
Related Concept Videos
Cardiopulmonary Resuscitation III: AED Use
Acute Coronary Syndrome III: Diagnostic Studies