The WATCHMAN Left Atrial Appendage Closure Device for Atrial Fibrillation
Published on: February 28, 2012
Pulmonary Artery Perforation by Atrial Appendage Closure Device and the Contribution of Auricular Contraction to Its
Teresa Bretones-Pino1, Ricardo Rivera-López1, Alejandro Carrero-Castaño2
1Department of Interventional Cardiology, Virgen de las Nieves University Hospital, Granada, Spain.
Abstract:
We report the case of an 83-year-old man with paroxysmal atrial fibrillation who underwent successful percutaneous left atrial appendage closure with the LAmbre device, being in sinus rhythm at implantation. Suddenly, the patient experienced cardiac tamponade and died within a few minutes. Autopsy revealed a slight protrusion of the LAmbre device into the atrial appendage wall, and pulmonary artery laceration. This is the first published report of pulmonary artery perforation by the LAmbre device. This case highlights the need for a detailed imaging study before this procedure is performed, to assess left atrial appendage movement/contraction in patients in sinus rhythm.